HemosIL® AcuStar™ Anti-Cardiolipin IgM Controls
Transcription
HemosIL® AcuStar™ Anti-Cardiolipin IgM Controls
HemosIL® AcuStar™ Anti-Cardiolipin IgM Controls - 0009802108 Intended use For the quality control of the Anti-Cardiolipin IgM assay (PN 0009802008) performed on the ACLTM AcuStar. Summary and principle Anti-cardiolipin (aCL) antibodies belong to a heterogeneous family of antiphospholipid (aPL) antibodies, which are autoantibodies directed against anionic phospholipids or protein-phospholipid complexes. Persistently elevated levels of aPL antibodies are associated with an increased risk for vascular thrombosis and obstetrical complications. This association is known as Antiphospholipid Syndrome (APS), classification proposed by Harris in 19871. Assays for the determination of aCL IgM and IgG antibodies, anti-ß2GPI IgG and IgM antibodies and Lupus Anticoagulant antibodies are the aPL assays included in the revised classification criteria determined by the International Committee for the diagnosis of APS in the meeting held in 2006 in Sydney (Australia)2, 3. The Low and High Anti-Cardiolipin (aCL) IgM Controls are prepared by means of a dedicated process and contain different concentrations of human aCL IgM antibodies. Low Anti-Cardiolipin IgM Control: Control intended for the assessment of precision and accuracy of the assay at normal or around cut-off aCL IgM levels. High Anti-Cardiolipin IgM Control: Control intended for the assessment of precision and accuracy of the assay at abnormal aCL IgM levels. Use of both controls is recommended for a complete quality control program. ENGLISH - Insert revision 10/2013 PRECAUTIONS AND WARNINGS: The human derived material in this product was tested by FDA approved methods and found nonreactive for Hepatitis B Surface Antigen (HBsAg), Anti-HCV and HIV 1/2 antibodies. Handle as if potentially infectious4. Hazard class: None Risk phrases: None Safety phrases: None This product is For in vitro Diagnostic Use. Preparation The aCL IgM Controls are liquid and must be mixed by gentle inversion several times before use to assure homogeneity. Avoid foam formation. Note: Bubbles on top of the liquids may interfere with the instruments liquid sensors. Controls storage and stability Unopened controls are stable until the expiration date shown on the tube label when stored at 2-8°C. Stability after opening: 10 hours continuously on-board the ACL AcuStar System (in open barcoded tube) or 4 days when placed twice daily on-board (in open barcoded tube) for 45 minutes and then recapped and placed at 2-8ºC. Controls can be frozen at -20ºC and thawed once. Do not refreeze. For optimal stability remove control from the system and store it at 2-8°C capped in the original vial. Composition Instrument/test procedures The AcuStar aCL IgM Controls kit consists of: L Low aCL IgM Control (Cat. No. 0009802109): 3 x 1 mL barcoded tubes of a solution with aCL IgM in a saline solution containing bovine fetal serum, stabilizers and preservative. H High aCL IgM Control (Cat. No. 0009802110): 3 x 1 mL barcoded tubes of a solution with aCL IgM in a saline solution containing bovine fetal serum, stabilizers and preservative. The aCL IgM Controls should be handled in the same manner as fresh citrated plasma or serum. Qualitätskontrollen für den Anti-Cardiolipin IgM Test (Art. Nr. 0009802008) auf dem ACLTM AcuStar. Testprinzip und Zusammenfassung Die anti-Cardiolipin (aCL) Antikörper gehören zur heterogenen Familie der antiPhospholipid (aPL) Antikörper, welche Autoantikörper darstellen, die gegen negativ geladene Phospholipide oder Phospholipid-Protein-Komplexe gerichtet sind. Anhaltend erhöhte Spiegel von aPL Antikörpern sind mit einem erhöhten Risiko für Gefäßthrombosen und Geburtskomplikationen assoziiert. Dieser Zusammenhang wird als Antiphospholipid Syndrom (APS) bezeichnet, die Klassifizierung wurde 1987 von Harris aufgestellt1. Teste für die Bestimmung von aCL IgM und IgG Antikörpern, anti-ß2 GPI IgG und IgM Antikörpern und Lupus Antikoagulanz Antikörpern sind die aPL Teste, die in den überarbeiteten Klassifizierungskriterien enthalten sind, die beim Treffen des International Committee for the diagnosis of APS, welches 2006 in Sydney (Australien) stattgefunden hat, erstellt wurden2,3. Die Anti-Cardiolipin (aCL) IgM Kontrollen Niedriger und Hoher Bereich wurden mittels eines speziellen Prozesses hergestellt und enthalten unterschiedliche Konzentrationen von humanen aCL IgM Antikörpern. Anti-Cardiolipin IgM Kontrolle Niedriger Bereich: Kontrolle zur Überprüfung der Präzision und Richtigkeit von aCL IgM im normalen Bereich oder im Bereich des Cut Off Wertes. Anti-Cardiolipin IgM Kontrolle Hoher Bereich: Kontrolle zur Überprüfung der Präzision und Richtigkeit von aCL IgM im abnormalen Bereich. Es wird empfohlen, die Qualität der Analyse mit beiden Kontrollbereichen zu überprüfen. Inhalt Die AcuStar aCL IgM Controls Packung enthält: L Low aCL IgM Control (Art. Nr. 0009802109): 3 x 1 mL barcodierte Röhrchen mit einer Lösung aus aCL IgM in Kochsalzlösung, die bovines fetales Serum, Stabilisatoren und Konservierungsmittel enthält. H High aCL IgM Control (Art. Nr. 0009802110): 3 x 1 mL barcodierte Röhrchen mit einer Lösung aus aCL IgM in Kochsalzlösung, die bovines fetales Serum, Stabilisatoren und Konservierungsmittel enthält. Para el control de calidad de la determinación de IgM anticardiolipina (Cat No. 0009802008) realizada con el ACLTM AcuStar. Sumario y principio Los anticuerpos anticardiolipina (aCL) pertenecen a la heterogénea familia de los anticuerpos antifosfolípidos (aPL), los cuales son autoanticuerpos contra fosfolípidos aniónicos o complejos proteína-fosfolípidos. Niveles de anticuerpos aPL persistentemente elevados se asocian con un aumento del riesgo de trombosis vascular y complicaciones obstétricas. Esta asociación se conoce como Síndrome Antifosfolípidos (APS), clasificación propuesta por Harris en 19871. Los ensayos para la determinación de anticuerpos IgG e IgM anti-ß2GPI, anticuerpos IgG e IgM anticardiolipina (aCL) y anticuerpos anticoagulantes lúpicos son los ensayos de aPL incluidos en los criterios de clasificación revisados y acordados en la convención del Comité Internacional para el diagnóstico de APS realizada en 2006 en Sydney (Australia)2,3. Los controles bajo y alto de IgM anticardiolipina (aCL) se han preparado aplicando un proceso específico y contienen diferentes concentraciones de anticuerpos IgM humanos anti-CL. Control bajo de IgM anti-Cardiolipina: Control utilizado para evaluar la precisión y la exactitud del ensayo en niveles normales o cerca del umbral de IgM aCL. Control alto de IgM anti-Cardiolipina: Control utilizado para evaluar la precisión y la exactitud del ensayo en niveles anormales de IgM aCL. Se recomienda el uso de ambos controles para un programa de control de calidad completo. Composición El kit AcuStar aCL IgM Controls está formado por: L Low aCL IgM Control (Cat. No. 0009802109): 3 x 1 mL de una solución de IgM aCL en solución salina que contiene suero fetal bovino, estabilizadores y conservante, en tubo con código de barras. H High aCL IgM Control (Cat. No. 0009802110): 3 x 1 mL de una solución de IgM aCL en solución salina que que contiene suero fetal bovino, estabilizadores y conservante, en tubo con código de barras. Limitations These products are designed as controls for monitoring the aCL IgM assay. These controls are subjected to the limitations of the assay system. Deviations may indicate possible problems with one or more components in the test system. Performance characteristics/expected values Refer to the AcuStar aCL IgM reagent insert for performance characteristics. The Low and High aCL IgM Controls concentration ranges were determined over multiple runs on the ACL AcuStar System using different lots of AcuStar aCL IgM reagents. The mean of the control range determined in each laboratory may vary due to the lot of reagent used. (*) Information required to create the controls in the ACL AcuStar. Enter only last 4 digits of the lot number The following is not supplied with the kit and must be purchased separately. AcuStar Anti-Cardiolipin IgM Cat. No. 0009802008 DEUTSCH - Packungsbeilage Version 10/2013 Zusätzliche Reagenzien und Kontrollmaterialien WARNUNG: Das verwendete Humanmaterial wurde mit FDA anerkannten Testmethoden als nicht reaktiv für Hepatitis-B-Antigen (HBsAg), Anti-HCV und HIV 1/2-Antiköper getestet. Als potentiell infektiöses Material behandeln.4 Gefahrenklasse: keine Risikoeinstufung: keine Sicherheitseinstufung: keine Dieses Produkt ist nur für die in vitro Diagnostik geeignet. Die folgenden Reagenzien sind nicht in der Packung enthalten und müssen zusätzlich bestellt werden: AcuStar Anti-Cardiolipin IgM Art. Nr. 0009802008 Standardisierung der Kontrollmaterialien Die aCL IgM Kontrollen sind flüssig und müssen durch mehrmaliges vorsichtiges Umschwenken vor der Verwendung gemischt werden, um die Homogenität sicherzustellen. Schaumbildung vermeiden. Hinweis: Auf der Flüssigkeitsoberfläche befindliche Luftblasen können den Flüssigkeitssensor des Gerätes beeinträchtigen. Die angegebenen Referenzwerte wurden bei mehreren Läufen am ACL AcuStar System mit spezifischen Reagenzchargen und gegen einen internen “HausStandard” ermittelt. aCL IgM Ergebnisse werden in U/mL (frei gewählte IgG Units) angegeben. Diese Einheiten wurden durch die Festlegung von 20 U/mL als Antwort auf die Obergrenze des Normalbereichs (ULNR) von 250 Citratplasmen einer Blutbank (99 % Perzentil) bestimmt. Die Einheiten des internen Hausstandards für die aCL IgM Antikörper wurden nach den Empfehlungen des Treffens des International Committee for the diagnosis of APS, welches in Sydney (Australien) stattgefunden hat2, mit dem monoklonalen Referenz Antikörper EY2C9 korreliert5. Lagerung und Haltbarkeit Einschränkungen Die ungeöffneten Kontrollen sind bei Lagerung zwischen 2-8°C bis zu dem auf dem Etikett des Röhrchens angegebenen Verfallsdatum haltbar. Haltbarkeit nach Öffnen: - 10 Stunden im ACL AcuStar System (im geöffneten barcodierten Röhrchen) - 4 Tage bei zweimal täglicher Platzierung im ACL AcuStar System (im geöffneten barcodierten Röhrchen) für 45 Minuten, anschließendem Verschluss und Lagerung bei 2-8°C - Kontrollen können bei -20°C eingefroren und einmalig aufgetaut werden; nicht erneut einfrieren Für eine optimale Haltbarkeit sollten die Kontrollen aus dem Gerät entnommen und im Kühlschrank bei 2-8°C im verschlossenen barcodierten Plastikröhrchen aufbewahrt werden. Dieses Produkt ist als Kontrolle zur Überprüfung des aCL IgM Testes vorgesehen. Die Kontrollen unterliegen den Einschränkungen durch das Testsystem. Abweichungen weisen auf mögliche Probleme von Komponenten des Testsystems hin. Herstellung Bestimmungsansatz Die aCL IgM Kontrollen sollten wie frisches Citratplasma oder Serum behandelt werden. Anti-Cardiolipin IgM Controls - 0009802108 Aplicación The reported values were determined over multiple runs on the ACL AcuStar System using specific lots of reagents and against an internal House Standard. aCL IgM results are reported in U/mL (IgM Arbitrary Units). These units have been established assigning 20 U/mL to the response of the upper limit of normality range (ULNR) of 250 citrated plasmas from a blood bank (99th percentile). Following the recommendations of the International Committee for the diagnosis of APS in the meeting held in Sydney2, the units of the internal House Standard for the aCL IgM antibodies have been correlated with the reference monoclonal antibody EY2C95. Additional reagents Anti-Cardiolipin IgM Controls - 0009802108 Verwendung Traceability of control materials Testcharakteristik und Referenzbereiche Die Informationen zur Testcharakteristik sind der Packungsbeilage des AcuStar aCL IgM Reagenzes zu entnehmen. Die Referenzbereiche der aCL IgM Kontrollen Niedriger und Hoher Bereich wurden bei mehreren Läufen am ACL AcuStar System mit verschiedenen Reagenzchargen des AcuStar aCL IgM Reagenzes bestimmt. Der in verschiedenen Laboren gefundene Mittelwert des Kontrollbereichs kann in Abhängigkeit der verwendeten Reagenzcharge variieren. (*) Information notwendig um Kontrollen im ACL AcuStar anlegen zu können. Nur die 4 letzten Ziffern der Chargennummer eingeben. ESPAÑOL - Revisión Folleto 10/2013 Reactivos adicionales PRECAUCIONES Y ADVERTENCIAS: El material de origen humano utilizado en este producto ha sido analizado por métodos aprobados por la FDA y se ha encontrado negativo para el antígeno de superficie de la hepatitis B (HBsAg) y para anticuerpos anti-HCV y HIV 1/2. Manejarlo con precaución como si fuese potencialmente infeccioso4. Indicaciones de peligro: ninguna Frases de riesgo: ninguna Frases de seguridad: ninguna Este producto es sólo para el diagnóstico in vitro. Los siguientes reactivos no se incluyen en el kit y deben adquirirse por separado. AcuStar Anti-Cardiolipin IgM Cat. No. 0009802008 Trazabilidad de los controles Preparación Los controles de IgM aCL son líquidos y se deben mezclar suavemente por inversión varias veces antes de usar para asegurar su homogeneidad. Evitar la formación de espuma. Nota: La presencia de burbujas en la capa superior de los líquidos puede interferir con los detectores de nivel de líquidos del instrumento. Conservación y estabilidad Los controles que no hayan sido abiertos son estables hasta la fecha de caducidad indicada en la etiqueta del tubo, si se conservan a 2-8°C. Estabilidad después de abrir: 10 horas seguidas a bordo del instrumento ACL AcuStar (en tubo abierto con código de barras) o 4 días cuando se coloca dos veces al día en el instrumento (en tubo abierto con código de barras) durante 45 minutos y después se vuelve a tapar y guardar a 2-8ºC. Los controles se pueden congelar a -20ºC y descongelar 1 vez. No se deben volver a congelar. Para una estabilidad óptima, retirar los controles del sistema y conservarlos a 2-8°C en los tubos originales con código de barras y tapón. Procedimiento del instrumento/ensayo Los controles de IgM aCL deben ser tratados de la misma manera que las muestras frescas de plasma citratado o suero. Los valores indicados se establecieron mediante múltiples determinaciones con el sistema ACL AcuStar utilizando lotes específicos de reactivos y se compararon con un estándar interno. Los resultados de IgM aCL se expresan en U/mL (Unidades Arbitrarias de IgM). Estas unidades se establecieron asignando 20 U/mL a la respuesta del limite superior del rango de normalidad (LSRN) de 250 plasmas citratados de banco de sangre (percentil 99). De acuerdo con las recomendaciones de la convención del Comité Internacional para el diagnóstico del APS realizada en Sydney2, las unidades del estándar interno para los anticuerpos IgM aCL fueron referenciadas al anticuerpo quimérico EY2C95. Limitaciones Estos productos han sido diseñados como controles para la monitorización del ensayo de IgM aCL. Estos controles están sujetos a las limitaciones del sistema del ensayo. Las desviaciones pueden indicar posibles problemas con uno o más componentes del sistema de análisis. Características técnicas/valores esperados Consultar el folleto del reactivo AcuStar aCL IgM para obtener información sobre las características técnicas. Los rangos de aceptación de los controles bajo y alto de IgM aCL se establecieron mediante múltiples determinaciones con el sistema ACL AcuStar utilizando lotes específicos de reactivos AcuStar aCL IgM. La media del rango de control establecida en cada laboratorio puede variar en función del lote de reactivo utilizado. (*)Información requerida para crear los controles en el ACL AcuStar. Entrar sólo los 4 últimos dígitos del número de lote. Bibliography / Literatur / Bibliografía / Bibliographie / Bibliografia /Bibliografia 1.Harris EN. Syndrome in the black swan, Br.J. Rheumatol. 1987; 26 324-326 2.Miyakis S, Lockshin MD, Atsumi T, Branch DW, Brey RL, Cervera R, Derksen RH, de Groot PG, Koike T, Meroni PL, Reber G, Shoenfeld Y, Tincani A, Vlachoyiannopoulos PG, Krilis SA. International consensus statement on an update of the classification criteria for definite antiphospholipid syndrome (APS), J Thromb Haemost. 2006;4(2): 295-306 3.Swadoêba J, Iwaniec T, Szczeklik A, MusiaÞ J. Revised classification criteria for antiphospholipid syndrome and the thrombotic risk in patients with autoimmune diseases, J Thromb Haemost. 2007;5: 1883-1889 4.Richmond JY, McKinney RW eds. Biosafety in Microbiological and Biomedical Laboratories, U.S. Dept. of Health and Human Services, Public Health Service, 4th Edition, 1999 5.Ichicawa K, Tsutsumi A, Atsumi T, Matsuura E, Kobayashi S, Hughes GR, Khamashta MA, Koike T. .A chimeric antibody with the human γ1 constant region as a putative standard for assays to detect IgG ß2-Glycoprotein-I-dependent anticardiolipin and anti-ß2-Glycoprotein I antibodies Arthritis and Rheumatism. 1999; 42 (11): 2461-2470 ACL and AcuStar are trademarks of Instrumentation Laboratory. ©2008 Instrumentation Laboratory. Issued October 2013 Symbols used / Verwendete Symbole / Símbolos utilizados / Symboles utilisés / Simboli impiegati / Símbolos utilizados CONTROL NAME NAME TARGET DOSE IL Control name / Namen steuern / Nombre del control / Contrôle / Controllare il nome / Nome do Controlo Name / Name / Nombre / Nom / Nome / Nome Target dose / Ziel-Dosis / Valor de referencia / Valeur médiane / Dose dell'obiettivo / Dose Alvo ACCEPTANCE RANGE Acceptance range / Referenzbereich / Rango esperado / Limites d’acceptabilité / Intervallo di accettabilità / Intervalo de aceitação UNITS Measurement units / Messeinheiten / Unidades de medida / Unités de mesure / Unità di misura / Unidades de Medida In vitro diagnostic medical device In-vitro Diagnostikum De uso diagnóstico in vitro Dispositif mèdical de diagnostic in vitro Per uso diagnostico in vitro Dispositivo médico para utilização em diagnóstico in vitro Batch code Chargen-Bezeichnung Identificación número de lote Désignation du lot Numero del lotto Número de lote Use by Verwendbar bis Caducidad Utilisable jusqu’à Da utilizzare prima del Data límite de utilização Temperature limitation Festgelegte Temperatur Temperatura de Almacenamiento Températures limites de conservation Limiti di temperatura Límite de temperatura Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) Consult instructions for use Beilage beachten Consultar la metódica Lire le mode d’emploi Vedere istruzioni per l’uso Consultar as instruções de utilização Control Kontrollen Control Contrôle Controllo Controlo Biological risks Biologisches Risiko Riesgo biológico Risque biologique Rischio biologico Risco biológico Manufacturer Hergestellt von Fabricado por Fabricant Prodotto da Fabricado por Authorised representative Bevollmächtigter Representante autorizado Mandataire Rappresentanza autorizzata Representante autorizado 3800-3967 R5 10/2013 HemosIL® AcuStar™ Anti-Cardiolipin IgM Controls - 0009802108 Utilisation Pour le contrôle de qualité du dosage des Anti-Cardiolipine IgM (Réf. 0009802008) réalisé sur l’analyseur ACLTM AcuStar. Principe Les anticorps anti-Cardiolipines (aCL) font partie d’une famille hétérogène d’anticorps anti-phospholipides (aPL), qui sont des auto-anticorps dirigés contre les phospholipides anioniques ou les complexes phospholipoprotéiques. La présence persistante de taux élevés d’aPL est associée à un risque accru de survenues de thromboses veineuses et de complications obstétricales. Cette association est connue comme le Syndrome des Anti Phospholipides (SAPL) en regard de la classification proposée par Harris en 19871. La détermination des anticorps anticardiolipines (aCL) IgM et IgG, des anticorps anti-ß2GPI de types IgG et IgM et des anticorps anticoagulants lupiques sont les tests aPL faisant partie de la révision des critères de classification établis lors de la réunion du Comité International pour le Diagnostic de l’APS qui s’est tenue à Sydney (Australie) en 20062, 3. Les Contrôles Anti-Cardiolipines (aCL) IgM Bas et Haut sont préparés au moyen d’un procédé spécifique. Ils contiennent différentes concentrations d’anticorps aCL IgM humaines. Contrôle Anti-Cardiolipines IgM Bas (Low Anti-Cardiolipin IgM Control): Contrôle à utiliser pour l’estimation de la précision et de l’exactitude dans le dosage des aCL IgM à un taux normal ou proche du cut-off. Contrôle Anti-Cardiolipines IgM Haut (High Anti-Cardiolipin IgM Control): Contrôle à utiliser pour l’estimation de la précision et de l’exactitude dans le dosage des aCL IgM à un taux anormal. L’utilisation de chacun des contrôles est recommandée pour un programme de contrôle de qualité complet. Composition La trousse AcuStar aCL IgM Controls (Contrôles aCL IgM) contient: L Low aCL IgM Control (Réf. . 0009802109): 3 tubes codes barrés de 1 ml d’une solution d’aCL IgM en solution saline, contenant de l’albumine bovine foetale, des stabilisants et un conservateur. H High aCL IgM Control (Réf. 0009802110): 3 tubes codes barrés de 1 ml d’une solution d’aCL IgM en solution saline, contenant de l’albumine bovine foetale, des stabilisants et un conservateur. FRANÇAIS - Révision de la notice 10/2013 PRECAUTIONS: Ces réactifs contiennent des produits d’origine humaine. Ils ont été trouvés négatifs pour les anticorps anti VIH 1/2, anti VHC et l’antigène de surface de l’hépatite B (AgHBs), en utilisant des trousses de dépistage de 3ème génération ou des méthodes approuvées par la FDA. Cependant, aucune technique ne permettant d’assurer l’absence totale du virus HIV ou de l’hépatite B ou de tout autre agent infectieux, ces réactifs sont à manipuler comme s’ils étaient potentiellement infectieux.4 Classification risque: Aucune Phrases risque: Aucune Phrases sécurité: Aucune Ce produit est à usage diagnostique in vitro Préparation Les contrôles aCL IgM sont liquides et doivent être mélangés plusieurs fois par inversion avant utilisation afin d’assurer une parfaite homogénéité. Eviter la formation de mousse. Note: La présence de bulles d’air en surface du liquide peut entraîner des interférences avec les détecteurs de fluides des instruments. Eviter la formation de mousse Conservation et stabilité des contrôles Conservés à 2-8°C, les contrôles non ouverts sont stables jusqu’à la date de péremption indiquée sur le tube. La stabilité du réactif après ouverture est de : 10 heures à bord de l’analyseur ACL AcuStar de façon permanente (dans le tube code barré ouvert) ou 4 jours lorsque le tube code barré ouvert est placé 2 fois par jour à bord de l’analyseur ACL AcuStar pendant 45 minutes puis rebouché et ensuite conservé à 2-8°C. Les contrôles peuvent être congelés à -20°C et décongelés une seule fois. Ne pas recongeler. Pour une stabilité optimale, retirer le contrôle de l’analyseur et le conserver à 2-8°C dans le tube code barré bouché. Procédures de test/Instrument Per il controllo di qualità del test Anti-Cardiolipina IgM (PN 000982008) eseguito su ACLTM AcuStar. Principio del metodo Gli anticorpi anti-cardiolipina (aCL) appartengono alla famiglia eterogenea degli anticorpi antifosfolipidi (aPL), che sono autoanticorpi diretti contro i fosfolipidi anionici o i complessi tra proteine e fosfolipidi. Livelli elevati persistenti di aPL sono associati ad un aumento del rischio di trombosi vascolari e complicanze ostetriche. Questa associazione è nota come sindrome da antifosfolipidi (APS), classificazione proposta da Harris nel 19871. La determinazione degli anticorpi IgM e IgG aCL, degli anticorpi IgG e IgM anti-ß2 Glicoproteina-I (anti-ß2GPI) e degli anticorpi Lupus Anticoagulant, sono i test aPL definiti negli ultimi criteri di classificazione indicati nel convegno dall’International Committe per la diagnosi dell’APS che ha avuto luogo a Sydney (Australia)2,3 nel 2006. I Controlli Anti-Cardiolipina (aCL) IgM Basso ed Alto sono preparati attraverso un processo dedicato e contengono concentrazioni diverse di anticorpi IgM aCL umano. Low Anti-Cardiolipin IgM Control: Controllo per la stima della precisione e dell’accuratezza del test degli IgM aCL a livelli normali o vicino al cut-off. High Anti-Cardiolipin IgM Control: Controllo per la stima della precisione e dell’accuratezza del test degli IgM aCL a livelli anormali. Per un programma di controllo di qualità completo si raccomanda l’uso di entrambi i controlli. Composizione Il kit AcuStar Anti-Cardiolipin IgM Controls è composto da: L Low aCL IgM Control (Nr. Cat. 0009802109): 3 provette di plastica con barcode da 1 mL di una soluzione salina di IgM aCL con siero fetale bovino, stabilizzanti e conservanti. Para o controlo de qualidade do ensaio Anti-Cardiolipin IgM (PN 0009802008) realizado no ACLTM AcuStar. Resumo e princípio Os anticorpos anti-cardiolipina (aCL) pertencem à família heterogénea dos anticorpos antifosfolípidos (aPL), que são autoanticorpos dirigidos contra fosfolípidos aniónicos ou complexos proteína-fosfolípidos. Níveis persistentemente elevados de anticorpos antifosfolípidos estão associados com um risco aumentado de tromboses vasculares ou complicações obstétricas. Esta associação é conhecida como Síndrome Antifosfolípidico (APS), classificação proposta por Harris em 19871. Os ensaios para a determinação de anticorpos aCL IgM e IgG, anti-anticorpos ß2GPI IgM e IgG e anticorpos anticoagulantes Lupus são ensaios aPL incluídos nos critérios de classificação revistos determinados pelo International Committee for the diagnosis of APS celebrado em 2006 em Sydney (Austrália)2,3 Os controlos IgM Low e High Anti-Cardiolipin (aCL) são preparados através de um processo dedicado e contêm diferentes concentrações de anticorpos IgM anti-CL humanos. Low Anti-Cardiolipin IgM Control: Controlo destinado à avaliação da precisão e da exactidão do ensaio com níveis normais ou aproximados dos cut-off do aCL IgM. High Anti-Cardiolipin IgM Control: Controlo destinado à avaliação da precisão e da exactidão do ensaio com níveis anormais aCL IgM. Recomenda-se a utilização de ambos os controlos para um programa completo de controlo da qualidade. Composição O kit AcuStar aCL IgM Controls consiste em: L Low aCL IgM Control (Cat. No. 0009802109): 3 tubos x 1 mL com código de barras com soro humano diluído positivo para aCL IgM, com soro bovino fetal, soro fisiológico, estabilizantes e conservante. Preparazione I Controlli aCL IgM sono liquidi e devono essere miscelati gentilmente diverse volte per inversione prima dell’uso. Evitare la formazione di schiuma. Nota: La presenza di bolle può interferire nel corretto funzionamento dei sensori dei liquidi sugli strumenti. Conservazione e stabilità dei reagenti I controlli sigillati sono stabili a 2-8°C fino alla data di scadenza stampata sull’etichetta della provetta. Stabilità dopo apertura: 10 ore sui sistemi ACL AcuStar (nelle provetta con barcode aperta) o 4 giorni quando caricata a bordo per due volte (nelle provetta con barcode aperta) per 45 minuti e poi richiusa e conservata a 2-8°C. I controlli possono essere congelati a -20°C e riutilizzati una volta sola. Non ricongelare Metodo Dopo la ricostituzione i controlli del aCL IgM vanno trattati come normali campioni di plasma citratato o siero fresco. L Low ACL IgM Control CONTROL NAME H High ACL IgM Control NAME LOT (*) (*) Expiration (*) Les valeurs reportées dans les notices sont déterminées à partir d’essais multiples réalisés sur l’analyseur ACL AcuStar, en utilisant des lots spécifiques de réactifs, et ce, par rapport à un Standard Interne. Les résultats aCL IgM sont reportés en U/ ml (Unités arbitraires IgM). Ces unités ont été établies en assignant la valeur de 20 U/ml à la réponse correspondant à la limite haute de la normalité (ULNR) établie à partir de 250 plasmas citratés issus de donneurs sains d’une banque du sang (99ème percentile). Les unités du Standard Interne pour les anticorps aCL IgM ont été corrélées en regard de l’anticorps monoclonal de référence EY2C95, selon les recommandations issues de la réunion du Comité International pour le Diagnostic de l’APS qui s’est tenue à Sydney2. Limites Ces produits sont utilisés en tant que contrôles dans le cadre du suivi du dosage des aCL IgM. Ces contrôles sont soumis aux mêmes limitations que le système de dosage. Des déviations peuvent indiquer un éventuel problème avec un ou plusieurs composants du système de dosage. Caractéristiques et performances/valeurs attendues Se référer à la fiche technique du réactif AcuStar aCL IgM pour les caractéristiques et performances. Les valeurs de concentrations des contrôles aCL IgM Bas et Haut ont été déterminées au cours d’essais multiples sur l’analyseur ACL AcuStar en utilisant des lots différents de réactifs AcuStar aCL IgM. La moyenne des valeurs de contrôle déterminée dans chaque laboratoire peut varier en fonction du lot de réactif utilisé. (*) Information requise pour créer les contrôles dans le système ACL AcuStar. Entrer seulement les 4 derniers caractères du numéro de lot. Reagenti ausiliari e plasmi di controllo I seguenti reagenti non sono forniti insieme al kit e devono essere ordinati separatamente. AcuStar Anti-Cardiolipin IgM Cat. No. 0009802008 Calibratori e materiale di controllo I valori indicati sono stati assegnati eseguendo repliche successive sui sistemi ACL AcuStar con l’impiego di lotti specifici di reagente rispetto a uno Standard Interno. I risultati di aCL IgM sono riportati in U/ml (Unità Arbitrarie IgM). Queste unità sono state stabilite assegnando 20 U/mL al limite superiore dell’intervallo di normalità (ULNR) di 250 plasmi citratati da campioni di donatori (99imo percentile). I valori dello Standard Interno per gli anticorpi IgM aCL sono stati assegnati con l’anticorpo monoclonale EY2C95 di riferimento, seguendo le indicazioni stabilite dall’International Committe per la diagnosi dell’APS di Sydney2. Limitazioni Questi prodotti sono destinati come controlli per la verifica del test aCL IgM. Questi controlli sono soggetti alle limitazioni del sistema di analisi. Eventuali risultati anomali possono evidenziare possibili problemi a carico di uno o più componenti del sistema analitico. Prestazioni e valori attesi Fare riferimento all’inserto del kit AcuStar aCL IgM per la verifica delle prestazioni analitiche. Gli intervalli di accettabilità dei Controlli aCL IgM Livello Basso ed Alto sono stati assegnati eseguendo ripetuti test sui sistemi ACL AcuStar con l’impiego di lotti specifici di reagente AcuStar aCL IgM. Il valore medio dell’intervallo di accettabilità determinato in ogni laboratorio può variare in funzione del lotto di reattivo utilizzato. (*) Codice di riconoscimento per creare i controlli su ACL AcuStar. Inserire solo le ultime 4 cifre. PORTUGUÊS - Revisão do folheto 10/2013 H High aCL IgM Control (Cat. No. 0009802110): 3 tubos x 1 mL com código de barras com soro humano diluído positivo para aCL IgM, com soro bovino fetal, soro fisiológico, estabilizantes e conservante. PRECAUÇÕES E AVISOS: O material de origem humana foi analisado com métodos de teste aprovados pela FDA e foi não reactivo para o Antigénio de Superfície da Hepatite B (HBsAg), AntiHCV e anticorpos HIV 1/2. Manipule como potencialmente infeccioso.4 Classe de risco: nenhuma Frases de risco: nenhuma Frases de segurança: nenhuma Este produto é para uso em diagnóstico in vitro. Preparação Os aCL IgM Controls são líquidos e devem ser misturados por inversão suave várias vezes, antes de usar para assegurar homogeneidade. Evite a formação de espuma. Nota: As bolhas no cimo dos líquidos podem interferir com os sensores de líquidos dos instrumentos. Conservação dos controlos e estabilidade Os controlos selados são estáveis até ao fim do prazo de validade impresso no rótulo, desde que conservados entre 2-8°C. Estabilidade pós-abertura: 10 horas continuamente no Sistema ACL AcuStar (num tubo com código de barras aberto ou 4 dias quando colocado 2 vezes por dia no sistema (num tubo aberto com código de barras), durante 45 minutos e depois coloca-se novamente a tampa e é conservado entre 2-8ºC.Os controlos podem ser congelados a -20ºC e descongelados uma vez. Não volte a congelar. Para estabilidade óptima, remova o controlo do sistema e conserve-o entre 2-8°C com tampa, no frasco original. Instrumento/procedimentos do teste Os controlos aCL IgM devem ser manipulados do mesmo modo, que plasma citratado ou soro. CONTROL NAME Détermination des valeurs des calibrateurs et contrôles ITALIANO - Revisione dell’inserto 10/2013 H High aCL IgM Control (Nr. Cat. 0009802110): 3 provette di plastica con barcode da 1 mL di una soluzione salina di IgM aCL con siero fetale bovino, stabilizzanti e conservanti. ATTENZIONE: Questo prodotto contiene materiale di cui è stata verificata all’origine, con metodi raccomandati dalla FDA, l’assenza dell’antigene di superficie dell’epatite B (HBsAg) e degli anticorpi anti-HCV e anti-HIV 1/2. Trattare come potenzialmente infetto.4 Simbolo di pericolo: nessuno Frasi di rischio: nessuno Consigli di prudenza: nessuno Per uso diagnostico in vitro. Anti-Cardiolipin IgM Controls - 0009802108 Utilização prevista Non fournis avec la trousse, ils doivent faire l’objet d’une commande séparée. AcuStar Anti-Cardiolipin IgM Réf. 0009802008 Les contrôles aCL IgM pourront être manipulés de la même manière qu’un plasma frais citraté ou un sérum. Anti-Cardiolipin IgM Controls - 0009802108 Utilizzo Réactifs auxiliaires et plasmas de contrôle TARGET DOSE ACCEPTANCE RANGE ACL AcuStar (*) ACL AcuStar ACLML Reagentes adicionais O seguinte reagente não é fornecido com o kit e deve ser adquirido separadamente. AcuStar Anti-Cardiolipin IgM Cat. No. 0009802008 Rastreabilidade dos materiais de controlo Os valores notificados foram determinados ao longo de múltiplas análises no Sistema ACL AcuStar utilizando lotes específicos de reagentes e contra um Padrão interno. Os resultados aCL IgM são notificados em U/mL (Unidades arbitrárias IgM). Estas unidades foram estabelecidas, atribuindo 20 U/ml à resposta do limite superior do intervalo normal (ULNR) de 250 plasmas citratados de um banco de sangue (percentil 99). Seguindo as recomendações do Encontro do International Committee for the diagnosis of APS em Sydney2 as unidades do Padrão interno para os anticorpos aCL IgM foram correlacionados com o anticorpo monoclonal de referência EY2C95. Limitações Estes produtos são concebidos como controlos para a monitorização do ensaio aCL IgM. Estes controlos estão sujeitos às limitações do sistema de ensaio. Os desvios podem indicar possíveis problemas com um ou mais componentes do sistema de teste. Características de desempenho/valores esperados Consulte o folheto informativo do reagente AcuStar aCL IgM para obter as características de desempenho. Os intervalos das concentrações dos controlos Low e High aCL IgM foram determinados ao longo de múltiplas análises no Sistema ACL AcuStar com diferentes lotes de reagentes AcuStar aCL IgM. A média do intervalo do controlo determinado em cada laboratório pode variar, devido ao lote de reagente utilizado. (*)informação necessária para criar os controlos no ACL AcuStar. Insira apenas os ultimos 4 dígitos do nº de lote. UNITS U/mL NAME LOT (*) (*) Expiration (*) TARGET DOSE ACCEPTANCE RANGE ACL AcuStar (*) ACL AcuStar ACLMH UNITS U/mL (*) Information required to create the controls in the ACL AcuStar Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) 3800-3967 R5 10/2013 Anti-Cardiolipin IgM Controls0009802108 Printed Insert Sheet: Revision: Issued: C.O.: 3800-3967 R5 10/2013 438050 LANGUAGES ENGLISH GERMAN SPANISH FRENCH ITALIAN PORTUGUESE TECHNICAL SPECS PAPER: SIZE: PRINT: PRINT COLOR: White paper, 50-60 g/m2 weight. 297 x 420 mm (DIN A3). Front/Back. Front - Top band Green Pantone 382, all remaining type in black. Back - All type in black. Instrumentation Laboratory Company - Bedford, MA 01730-2443 (USA) Instrumentation Laboratory SpA - V.le Monza 338 - 20128 Milano (Italy) 3800-3967 R5 10/2013